Laboratory fume hoods in Saudi Arabia are ventilated enclosures that keep chemical vapours and dust away from people working in petrochemical, university, hospital, food and mining labs. This page covers the types Saudi labs specify, how containment is tested under EN 14175, ASHRAE 110 and the ductless standards as of 2026, how Saudi heat, dust and humidity affect exhaust and make-up air, and how to compare Saudi suppliers.
0 suppliers found
Skip the search. Post your requirement and we'll connect you with verified fume hoods suppliers who quote for your order directly.
Tell us about fume hoods in Saudi Arabia and receive competitive quotes from verified suppliers — free, fast, and no obligation.
Saudi laboratories mostly specify ducted fume hoods connected to a roof exhaust, with variable air volume hoods in larger buildings, acid-resistant hoods for digestion work and ductless filtration hoods only for small, known chemical loads. They are part of the laboratory equipment family in Saudi Arabia, and on most projects they are bought with the furniture and the ventilation system rather than on their own.
| Hood type | Where Saudi labs use it | What to specify |
|---|---|---|
| Ducted bypass hood (constant volume) | University teaching and research labs in Riyadh, Jeddah and Dhahran | Width and depth, sash arrangement, bypass, airflow indicator with audible alarm |
| VAV ducted hood | New research buildings and large QC labs where cooling energy is high | Controller response, sash sensor, automatic sash closer, test to EN 14175-6 |
| Acid digestion and high heat hood | Petrochemical, mining and environmental labs in Jubail, Yanbu and at mine sites | Polypropylene or ceramic liner, EN 14175-7 design, wash-down for perchloric acid |
| Walk-in hood | Pilot plants, large distillation and catalyst rigs | Clear internal height, spill sill, sash design for loading |
| Ductless filtration hood | Small volumes of listed chemicals where a duct cannot be routed | Class under NF X15-211 or EN 17242, carbon filter type, H14 particulate filter, fault alarm, chemical listing |
Hoods sit on base cabinets and share worktops and services with the lab furniture, so agree interfaces early: sink and gas outlets, cabinet ventilation for solvent storage, and who connects the duct. Solvent handling for instruments such as spectrophotometers belongs inside the hood. For microorganisms, buy a biosafety cabinet instead; a fume hood does not protect the sample.
A fume hood is shown to be safe by a containment test with tracer gas, done once on the design and again after installation. Face velocity readings are useful for routine checks but do not prove containment.
The European route is EN 14175: Part 3 type tests the model, Part 4 tests it on site, Part 6 covers VAV hoods and Part 7 high heat and acid hoods. The American route is ASHRAE 110-2016, which ASHRAE Journal describes as releasing SF6 tracer gas at 4 litres per minute into the hood and measuring it at a mannequin's breathing zone. Esco summarises the 2016 changes, including thermal anemometers and more readings per grid point, and the three ratings: AM in the factory, AI after installation and AU with the lab's own equipment. ANSI/ASSP Z9.5 adds design and operating guidance; Labconco cites its advice that exhaust volume should change by less than five percent as the sash moves on a bypass hood.
Ductless hoods follow AFNOR NF X15-211:2009, which has two classes, and the newer EN 17242:2025. Erlab's datasheet lists HEPA H14 particulate filters and an automatic alarm for filtration faults on its Class 1 configurations. Whatever the type, require the type test report with the bid and an on-site test before acceptance.
In Saudi Arabia the exhaust and make-up air system decides whether a hood passes its test and what it costs to run. Very hot summer air, fine dust and coastal humidity all act on that system.
As of 2026, fume hoods are not medical devices and have no dedicated product registration in Saudi Arabia; buyers specify and accept them on EN 14175 or ASHRAE 110 test results, while electrical parts such as fans and controllers may fall under SASO regulations.
| Standard or rule | Scope | Body |
|---|---|---|
| EN 14175-2, -3, -4, -6, -7 | Fume cupboard requirements, type tests, on-site tests, VAV, high heat and acid | CEN |
| ASHRAE 110-2016 | Method of testing laboratory fume hood performance (AM, AI, AU) | ASHRAE |
| ANSI/ASSP Z9.5 | Laboratory ventilation | ANSI and ASSP |
| NF X15-211:2009; EN 17242:2025 | Ductless recirculating filtration fume hoods | AFNOR; CEN |
| SASO technical regulations and SABER, where they apply | Electrical components supplied with the hood | SASO |
| ISO/IEC 17025 and ISO 15189 | Accredited labs, whose safety and environment controls include hoods | Saudi Accreditation Center (SAAC) |
The SASO and SABER guide explains how to check whether an electrical component needs a certificate before shipment. Large industrial operators often add their own laboratory and ventilation specifications for projects on their sites, so ask the project engineer for them before tendering.
Fume hoods in Saudi Arabia are supplied by lab furniture contractors and scientific distributors in Riyadh, Jeddah and Dammam, usually within a turnkey lab package; public labs buy through government tenders with local content requirements. Hold one party responsible for the on-site containment test.
Makers carried here include BIOBASE (ducted fume hoods within its air protection range) and Infitek (fume hoods in its laboratory furniture line). Ask for the type test report and the standard, liner chemical resistance data, monitor and alarm details, and a clear split of scope with the HVAC contractor. Labs with centrifuges and autoclaves in the same room should coordinate locations, since autoclaves release heat and steam and nearby centrifuges should not sit in the hood's airflow path.
Price depends on size, liner material, VAV controls, automatic sashes and wash-down systems, and on the exhaust and make-up air works, which often cost more than the hoods. Hoods are built to the lab layout and installed once services are ready, then tested, so their delivery follows the construction programme. To get comparable offers from Saudi suppliers, post your requirement with the hood schedule and chemicals used.